  • Abatement Technologies HEPA-CARE UV800F Germicidal UV Module

    The UV800F module installs easily onto the HC800F in just a few minutes. Simply slide it into the offset tabs of the clean air exhaust grill on top of the HC800F, and secure it in place with the included fasteners. The short UV800F power cord plugs directly into a GFCI receptacle on the side of the HC800F unit, so no separate wall receptacle is required. Once installed, there is typically no need to remove it as the UV800F moves with the HC800F.

    Built-in performance features of the UV800F Germicidal Module include:

    • High-Output - Four T6-type, ultra-high output 39-watt UVGI lamps (156 watts total) generate more than 17,000 µw/cm² of UV energy at a distance of 3 inches from the lamps
    • Long Life - Lamps rated for 9,000 hours of continuous operation
    • Maximum UV Exposure - Internal air baffles maximize UVGI exposure by slowing down the airflow and funneling it closer to the lamps
    • Easy Lamp Replacement - Easy room-side access facilitates lamp replacement

    Safety features of the UV800F include:

    • Complete Lamp Shielding - Prevents any direct UV exposure during operation
    • Safety Interlocks - Instantly deactivate lamps if the lamp compartment is opened or airflow is interrupted
    • Lamp On/Off Switch - Turns off the lamp so HC800F HEPA unit can run with or without UV
    • NRTL-Certified - NRTL-certified to conform with applicable UL & CSA safety standards

    Multiple Uses within Hospitals and Healthcare Facilities

    The combination of HEPA filtration, negative pressure, and UVGI disinfection is ideal for isolation rooms, ER, and patient isolation when responding to potential bioterrorism incidents or infectious illness outbreaks and during normal use. This combination is also perfect for disinfecting room air between patients. Uses for the HC800F/UV800F include:

    • Airborne Infectious Isolation Rooms - The most common uses are for irradiation of HEPA filtered air before it is exhausted from the room and disinfection of room air between patients
    • Protective Environments & Operating Rooms - The UV800F can provide added protection from airborne pathogens for immune-compromised and other sensitive patients
    • Continuous Air Cleaning & Recirculation - The air in almost any room or area can be cleaned, deodorized, and recirculated back into the same area. Common areas of use include patient waiting rooms, bronchoscopy, pediatrics, autopsy, and outpatient clinics

    Lamp Characteristics

    Lamp Base & Type: T6 19mm Four Pin w/ Ridge

    Lamp Bulb Length: 273mm ARC 169mm

    Bulb Diameter: 19mm

    Operating Current: 865mA

    Lamp Voltage: 26V

    Lamp Wattage: 23W

    UV Output

    Output UVC Watts: 6.8

    Output: µ/cm2 @ 1m 65

    Rated Life (hours): 12000

    % Output @ End of Life: 80.00%